中文

News

Illustrated Guide to On-Site Sensor Installation for Blasting Vibration Meters

Zhongke Instruments · News2014-01-25 11:13:41

Home/News/Illustrated Guide to On-Site Sensor Installation for Blasting Vibration Meters
Content

Illustrated Guide to On-Site Standalone Operation of the TC-4850 Blasting Vibration Meter

I.  Connecting the triaxial velocity sensor to the signal transmission cable

Key point: align the concave and convex parts

 

  Connecting the triaxial sensor to the vibration meter

II.  Sensor installation (gypsum used in this case)

1. Clean the point position with a brush to help the gypsum bond tightly to the ground surface

Key points:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2. Mix gypsum powder with water into a thick paste and spread evenly on the cleaned surface

 

 

 3. After confirming the level direction (in blasting, the sensor X direction usually points to the blast center), press the triaxial sensor flat with the palm, moving it side to side while pressing - this effectively prevents uneven bonding and air bubbles between the sensor and gypsum, improving the rigid connection between the sensor and the inspected surface

 

 

4. Add gypsum powder around the sensor to absorb moisture, accelerating gypsum setting

 

 

That is roughly all for installing blasting vibration meter sensor geophones.www.zkck.comZKCK is the first choice