Compliance Analysis of TC-4850 Series Blasting Vibration Meters with DL/T 5333-2021

Zhongke Instruments · News2026-08-19 09:37:59

Home/News/Compliance Analysis of TC-4850 Series Blasting Vibration Meters with DL/T 5333-2021
Content

1. Introduction

DL/T 5333-2021 "Safety Monitoring Regulation for Blasting in Hydropower and Water Conservancy Projects" is the core execution standard for blasting vibration monitoring in the hydropower and water conservancy field. It sets mandatory requirements for the technical parameters, metrological qualification and functional configuration of monitoring instruments. As mainstream equipment in China's engineering blasting monitoring field, the compliance of the ZKCK TC-4850 series blasting vibration meters has always been a key concern of engineering users.

This article performs a clause-by-clause benchmark analysis of the full TC-4850-3/6 and TC-4850N-3/6 product line against the requirements of DL/T 5333-2021, covering technical indicators, functional configuration and certification qualifications, providing a technical reference for equipment selection in blasting monitoring of hydropower projects.

2. Core Requirements of DL/T 5333-2021 for Monitoring Equipment

Chapter 3 "General Provisions" and Chapter 6 "Monitoring Instruments" of DL/T 5333-2021 contain detailed provisions on sensors, recording equipment and metrological calibration of blasting vibration test systems. The core mandatory requirements cover the following ten dimensions:

  1. The linear frequency band of the sensor must fully cover the measured vibration frequency

  2. Sampling frequency must be no less than 12 times the upper dominant vibration frequency of the measured signal

  3. The measurement amplitude range must satisfy the estimated vibration magnitude of the project

  4. Must possess a metrological qualification certificate and statutory metrology license mark

  5. Must support on-site display, parameter setting and waveform data storage

  6. Must have on-site data analysis capability and a computer communication interface

  7. Sensors and recording equipment must have a unique electronic identity identifier

  8. Monitoring points must use triaxial integrated sensors for vibration testing

  9. Environmental adaptability must satisfy a working temperature of -10°C to 50°C

  10. Instruments must be calibrated at the prescribed interval and remain within the validity period

3. Clause-by-Clause Compliance Comparison of the TC-4850 Series

The TC-4850-3/6 and TC-4850N-3/6 series blasting vibration meters share a unified hardware architecture and sensor system with identical core technical indicators; configuration differences exist only in channel count, storage capacity and network functions. The benchmark analysis of the full product line against the regulation follows:

Table

No.ClauseSpecific RequirementTC-4850 Series Technical ParametersCompliance
1Clause 6.0.2-1, Table 6.0.2Particle vibration velocity frequency band: far zone 2-500 Hz, near zone up to 500 HzStandard triaxial velocity sensor 5-500 Hz; main unit frequency response 0-10 kHz; optional low-frequency sensor (1-500 Hz) and digital low-frequency functionCompliant; low-frequency band expandable
2Clause 6.0.2-2Sampling frequency > 12x upper dominant frequency (500 Hz upper limit requires > 6,000 sps)Sampling rate 10 sps to 100 Ksps continuously adjustable; 10 Ksps commonly used in engineering✅ Compliant
3Clause 6.0.2-3Measurement amplitude range matches the project estimateVelocity range ±35 cm/s with adaptive range switching✅ Compliant
4Clause 6.0.2-5Possesses metrological qualification certificate and statutory metrology license markFull line holds CPA Pattern Approval Certificates of Measuring Instruments; statutory calibration certificates available✅ Compliant
5Clause 6.0.2-6On-site display, parameter input, waveform and data storageIndustrial-grade display and physical keys; on-site waveform/feature-value viewing; built-in solid-state storage✅ Compliant
6Clause 6.0.2-6Should have on-site data analysis and remote communication capabilityDedicated analysis software with FFT, filtering and Sadovsky regression; USB/485 communication; N series optional network transmission✅ Compliant
7Clause 6.0.2-7Sensors and recording equipment must have a unique electronic identity identifierBoth main unit and sensor carry unique factory serial numbers - traceable and identifiable✅ Compliant
8Clause 6.0.3-2Triaxial sensor installed at every monitoring pointStandard integrated XYZ triaxial velocity sensor with high orthogonality accuracy✅ Compliant
9Clause 3.0.8Ambient temperature range -10°C to 50°CWorking temperature -10°C to 75°C, humidity 0-95% RH✅ Compliant
10Clause 3.0.7Calibrated as prescribed and within the calibration validity periodSupports zero calibration and system calibration; periodic traceability at statutory metrology institutes✅ Compliant (performed by the user)

4. In-depth Analysis of Key Technical Points

4.1 Metrological Qualification: Policy Background of CPA Replacing CMC

According to the Metrology Law of the People's Republic of China (revised 2017), the former "Manufacturing License of Measuring Instruments (CMC)" system has been formally abolished; the current statutory metrological access qualification is thePattern Approval Certificate of Measuring Instruments (CPA). The two differ in regulatory focus: CMC examined enterprise production capability, while CPA focuses on the pattern compliance and metrological performance of the product itself, with higher legal effect and access threshold. It is the statutory basis for the legal manufacture, sale and engineering use of measuring instruments.

The full TC-4850 line has obtained CPA Pattern Approval Certificates, meeting the statutory metrological qualification requirements of DL/T 5333-2021. Before putting equipment into service, the user may send it to a statutory metrology institute for periodic calibration; with the calibration certificate it can be used for formal engineering monitoring.

4.2 Low-Frequency Response Capability and Extension Solutions

Table 6.0.2 of DL/T 5333-2021 sets the lower vibration frequency limit at 2 Hz for the far zone of cavern blasting and deep-hole blasting. The standard velocity sensor of the full TC-4850 line has a linear frequency response of 5-500 Hz. For 2-5 Hz low-frequency monitoring needs, two proven solutions are available:

Option 1: Optional low-frequency sensorA dedicated low-frequency velocity sensor covering 1-500 Hz can be paired, whose physical frequency response directly covers the 2 Hz lower limit - suitable for far-zone monitoring and high-yield blasting monitoring with strict low-frequency accuracy requirements.

Option 2: Enable the digital low-frequency functionThe instrument has a built-in digital low-frequency algorithm that extends the effective low-frequency bandwidth via digital signal processing, fully achieving 1-5 Hz low-frequency response characteristics without hardware replacement - suitable for routine engineering monitoring.

In actual engineering, the dominant blasting vibration frequency is mostly above 10 Hz and the energy share below 5 Hz is extremely low, so the standard sensor satisfies the blasting monitoring needs of the near-middle zones of the vast majority of hydropower projects.

4.3 Unique Electronic Identity Identifier

"Unique electronic identity identifier" is a mandatory requirement newly added in DL/T 5333-2021, intended to strengthen the traceability of monitoring equipment and ensure the authenticity of monitoring data. TC-4850 series main units and sensors are each assigned a unique device serial number and sensor number at the factory, with a complete product traceability system.

4.4 Extension Note on Acceleration Monitoring

If a project involves particle vibration acceleration monitoring, the acceleration measurement frequency upper limit in Table 6.0.2 reaches 3,000 Hz. The 0-10 kHz frequency response of the TC-4850 series main unit can meet the requirement, but an acceleration sensor must be additionally configured; the standard velocity sensor is not suitable for direct acceleration measurement.

5. Engineering Selection Recommendations

  1. Routine hydropower project blasting monitoring: The standard TC-4850-3/6 and TC-4850N-3/6 versions already satisfy the core requirements of DL/T 5333-2021, suitable for the vast majority of near-middle-zone blasting vibration monitoring scenarios.

  2. Far-zone low-frequency monitoring projects: The optional 1-500 Hz low-frequency sensor is recommended to ensure measurement accuracy in the 2 Hz low-frequency band and fully cover the frequency range required by the regulation.

  3. Network-based centralized monitoring projects: Prefer the TC-4850N series, whose network transmission enables remote centralized management of multiple monitoring points - fitting multi-point, long-period monitoring needs of large hydropower projects.

  4. Compliance documentation preparation: At purchase, request from the manufacturer copies of the CPA certificate and metrological calibration certificates of the corresponding model to ensure complete project acceptance documentation. Periodic calibration must be completed per the regulation before the equipment is put into service.

6. Conclusion

In summary, the TC-4850-3/6 and TC-4850N-3/6 series blasting vibration meters comply with the requirements of DL/T 5333-2021 "Safety Monitoring Regulation for Blasting in Hydropower and Water Conservancy Projects" in all core indicators - sampling rate, range accuracy, triaxial testing, on-site operation, metrological qualification and environmental adaptability. With the optional low-frequency sensor or the digital low-frequency function, the full frequency range specified by the regulation can be completely covered, satisfying the compliance and technical needs of blasting safety monitoring in all types of hydropower and water conservancy projects.